Operation of each of the 6 front panel keys may be set independently to the following settings:
Normal operation is the normal function of the key
Lock operation means that the key is locked and its normal operation is blocked
Immediate operation allows for the key function to operate without waiting for stable readings
The gross/net key (2) does not support immediate operation.
Example
Command
Answer
Description
S01;
Select unit 1
LBT?0;
1 CRLF
Zero key is currently set to normal operation
LBT0,0;
0 CRLF
Block operation of the zero key
TDD1;
0 CRLF
Save setting
21.3.26
LDN: Zero Calibration
Calibrate the zero dead weight of the scale. This is the new form of the LDW; command (see Section 21.3.27).
Because the C500 series has no concept of normal and direct calibration mode, it was necessary to introduce
a new command.
Note
There are two different methods of calibration. Either calibrate with standard
weight or direct input of mV/V.
General
No. of parameters
2
Save changes
With
TDD1
Increment Trade Counter
Yes
Parameter Details
Parameter
Description
Range
Default
1
Zero signal in deci-micro volts
per volt.
10000duV/V = 1000uV/V =
1 mV/V
-30000 .. 30000
-
2
Status (see table below)
0 .. 106
-
Calibration Status
Status value
Description
0
Calibration finished successfully
1
Calibration in process (busy)
101
Error zero too high (
>
3.0 mV/V) or too low (
<
-3.0 mV/V). Calibration aborted.
106
Other calibration error. Calibration aborted.
Example: Normal Calibration with Weight
The calibration process takes some time to complete. As a result it is necessary to monitor the calibration
process to determine when it is finished. To do this, issue a LDN? query.
138
C500-600-2.0.3
Summary of Contents for C500 series
Page 1: ...C500 Series Digital Indicator Reference Manual ...
Page 54: ...8 3 Custom Format Tokens 8 3 1 Formatting 49 C500 600 2 0 3 ...
Page 56: ...230 Send time in format hh mm ss 231 Send Date in format dd mm yyyy 51 C500 600 2 0 3 ...
Page 123: ...118 C500 600 2 0 3 ...
Page 182: ...177 C500 600 2 0 3 ...