91
Preparations
Inter-unit synchronization
Using two DA-21 units and the separately available dedicated cable, synchronized
operation with one unit functioning as master and one unit as slave is possible.
Important
Establish the same settings on both DA-21 units
(<Rec.Parameters> menu items). If settings are
different, synchronized operation is not possible.
Also, when the settings are changed after the
synchronization, it is necessary to synchronize
again.
Check the synchronization information on the
display (page 18) after operation of the syn-
chronization.
Note
The size of data fi les recorded simultaneously on two
synchronized DA-21 units will be different for the
master unit and the slave unit.
Synchronization procedure
1. Use the optional inter-unit sync cable CC-43 to link the Unit to Unit
connectors of the two DA-21 units that are to be synchronized.
Unit to Unit connector
Inter-unit sync cable CC-43