87
Calibration
Calibration Parameter
The Calibration Parameter is displayed on the “KL-04 - Calibration Parameter” window.
The explanations for each parameter are below.
Calibration:
Select a Calibration Parameter to show its setting details.
When a new Calibration Parameter Setting is required, enter the cali-
bration parameter in the space below “Calibration”.
Rev.:
The latest Revision Number (the number of changes) will be dis-
played.
The selected setting details for the Revision will be displayed when a
number is selected. The number is selected with the + – buttons.
Syringe
Syringe Volume (mL):
Set the capacity for the syringe installed on the unit.
The allowable range is 5 to 25 mL (1 mL step).
Important
Please be sure to set the installed syringe volume (the syringe
volume for standard installation is 25 mL). If set differently,
correct measurement may not be possible.