100
Performance-Test Result
KL-04 Performance-Test Data
Page 1 of 2
Performance Test:
#1: KL 04_Basic
Revision:
6
Creation Date:
2001 10 02 11:17:38
Operator:
RION’s Staffs (rion)
Instrument:
KL 04 No.00710003 (kl 04 0003)
RION
Operator
Manager
2001-12-18 13:02:20 setup@kl-04-3
Performance Test Parameter
Performance Test:
#1: KL 04_Basic (Rev. 12)
Aspiration Flow Rate:
25 mL/min
Measurement Volume:
5 mL
Number of Pre Measurements:
1
Number of Measurements:
3
Creation Date:
2001 10 02 09:48:42
Operator:
RION’s Staffs (rion)
Management Information:
File:
/home/kl04/share/perf/conf/1/12
Owner:
rion / kl04
Time Stamp:
2001 10 02 09:48:42
Fingerprint:
4d352d799f09ca19
Calibration Data
Calibration Data:
#1: KL 04_Basic (Rev. 8)
Aspiration Flow Rate:
25 mL/min
Creation Date:
2001 10 02 11:09:49
Operator:
RION’s Staffs (rion)
Management Information:
File:
/home/kl04/share/cal/data/1/8
Owner:
rion / kl04
Time Stamp:
2001 10 02 11:09:51
Fingerprint:
2129a70d3b83e7af
10.350µm
PASS
5.0µm
10.350µm
0.000µm
15.652µm
Test Method:
JP
Average Diameter of the PSL:
10.350µm
Subtract CV for the Resolution?
Yes
CV of the PSL:
1%
100% Cumulative Count Threshold:
5µm
Date:
2001 10 02 11:12:24
Operator:
RION’s Staffs (rion)
Concentration:
854 /mL [800 1200 /mL]
Sensor Resolution:
7.9% [max. 10%]
Threshold Accuracy:
0.5% [max. 5%]
Detail:
84% Cumulative Count Diameter:
9.972µm
50% Cumulative Count Diameter:
10.399µm
16% Cumulative Count Diameter:
11.224µm
Management Information:
File:
/home/kl04/share/perf/data/raw/10.35/138
Owner:
rion / kl04
Time Stamp:
2001 10 02 11:13:38
Fingerprint:
940f8220efb340c2