v
Caution
Types of sample fl uids
The unit is designed for measuring liquid-borne particles.
Never use the unit with the liquid types listed below,
to prevent false measurement and
the possibility of accidents and damage to the unit.
- Explosive liquid or mixtures containing explosive fl uid
- Hydrofl uoric acid or other fl uid which can cause corrosion of parts in contact with
the fl uid or mixtures containing such fl uid
- Liquid with a temperature of less than +15°C or more than +30°C
When using different types of sample fl uid where mixing can lead to reactions such as
thermal build-up, hardening, or particle generation, always perform thorough cleaning to
ensure complete replacement.
Weight
The weight of the unit is about 14 kg.
When transporting or otherwise moving the unit, use a suitable trolley or other implement
and ensure safety to prevent the risk of injury.
When setting the unit, hold the handle tight with one hand and with the other hand hold
the cut-out on the opposite side panel.
Summary of Contents for KL-04
Page 2: ......
Page 6: ...iv...
Page 10: ...viii...
Page 12: ...x...
Page 16: ...xiv...
Page 166: ...No 31947 06 11 Printed in Japan...