12
Preparations
Measurement with toxic or reactive gases
Before starting a measurement with toxic or reactive gases, be sure to perform
the following measures.
- Install and operate the system only as described in this manual.
- Connect a constant source of inert gas to the PURGE inlet connector
and connect the PURGE outlet connector to a suitable exhaust system.
- Before use, perform a leak test on the system and all piping including
the sample gas supply and exhaust piping, and the purge gas supply
and exhaust piping.
- Before introducing the sample gas, purge the piping with inert gas.
- Install gas sensors and suitable equipment to ensure that any gas leak
will be detected quickly and a suitable alarm will be activated.
- Be sure to have the unit serviced and checked regularly, at least once a
year. To arrange for servicing, contact the supplier.