17
Preparations
Turning the system on
Set the POWER switch of the KZ-50 and KS-93 to ON.
The power to the KS-93 will turn on in the initial condition.
Important
After turning on the power, wait at least 10 min-
utes before starting the measurement. This al-
lows the light source output to stabilize.
After turning the system off, wait at least 1 minute
before turning it on again. Otherwise the light
source may be damaged.
Initial condition
After power-on, the unit is in the condition as shown below. The initialization
command "C" (see page 31) also establishes this condition (except for the
communication settings).
- Light source is lit.
- Impossible to control light source on/off (local mode is set) (see page 32).
- Measurement is not performed.
- Alarm is released (terminals open, indicators off).
- Setting of transfer rate/node address setting switch bank is read.
(This does not apply when the initialization command "C" was
received.)
- Measurement data transfer mode is set to automatic send (see page 33).