34
Serial Interface
Measurement start/stop Command "Gn"
G0
Stops the measurement. If the command is sent while no
measurement is being carried out, the response message "R/ER3"
is returned (see page 44).
Note
In the mode for automatic sending of measurement
results (S0 mode), the KS-93 automatically transmits
measurement data when the unit receives a "G0" com-
mand.
G1
Starts the measurement. If the command is sent while measurement
cannot be carried out (for example because the light source is not
turned on), the measurement does not start and the response
message "R/ER3" is returned (see page 44).
G2
Cancels the measurement. In this case, measurement data are not
output.
Example:
X/G1
<EOL>