39
Serial Interface
Current error information message
The current error information message indicates information on the current
error status of the KS-93. The message is output in response to the current
error information request message "Q/K".
The following messages are used.
If two or more errors have occurred, the message for the error with the highest
priority is returned.
(Listed in order of priority)
K/COVER_OPEN
<EOL>
The cover of the unit is open (measurement
cannot be carried out).
K/LASER_OFF
<EOL>
The laser light source is turned off (measurement
cannot be carried out).
K/TEMP_NG
<EOL>
The light source temperature is out of range.
K/LASER_NG
<EOL>
The laser light source is not operating properly
(laser output level drop, etc.).
K/CELL_NG
<EOL>
There is a problem with the flow path (cell
contamination etc.).
K/LD_UNSTABLE
<EOL>
The laser output is unstable.
K/
<EOL>
No error has occurred.
_ indicates a space.
Note
Except for the "LASER_OFF" and "COVER_OPEN"
errors, measurement can be started and data will be
returned, but there might be a problem with data reli-
ability.