background image

 

Instrument Description 

 

1. 

Input for thermocouple         

probes 

2. 

LCD display

with 

Back-light

 

3. 

              key (Hold & 

Data) 

4. 

              key 

5. 

      key

power

 

6. 

External power supply   

(AC adapter) 

7. 

CAL calibration key   

8. 

  key (Back-light) 

9. 

              key 

10.  Signal output 
11.  Instrument model no.&   

S/N no. 

12.  Battery cover 

 

Specification of TK-62

 

 

 

※ 

Above data of “Accuracy” and “Sampling rate” exclude the error of temperature 
probes. 

※ 

The specification of TK-

62 will be the lasted version of Rixen, and it’s subject to 

change without notice. 

 

RS-232 Transmission Functions 

The RS-232 transmission port of TK-62 is single side data transmission, it 
functions with 3 signal composed grounding wires to receive and input. We 
suggest that you choose the original transmission cable or shorter than 10 
meters cable to connect PC and instrument. 

a. Transmission Status: 57600 / 8 / N / 1 
b. Transmission Content (8BIT): 

START 

TYPE 

MOD 

MSD 

LSD 

CHECK 

END 

A8H 

 

 

HI-DATA 

LO-DATA 

FFH 

8AH 

 

BIT [7:0] = Instrument Model 

                  TK-62.R=00H 

 

BIT [7:4] = Thermocouple Type 

                               
 

 
 
 

BIT [3:2] = Instrument Status 

 
 
 
 
 

BIT [1] = Power Model 

                          Automatic Power Off = 0 
                          Manual Power Off = 1 

 

BIT [0] = Power 

                          Enough Power = 0 
                          Low Power = 1 

c. 

Illustration of Connection:

 

 
 
 
 
 

Instrument Description 

Input 

Thermocouple K-TYPE (CA) 

Thermocouple R-TYPE 

(PR) 

Measuring Range 

- 200 ~ +1372

    (-328.0 ~ 

+ 1372

- 20 ~ +1772

    (- 4 ~ 

+3221

Resolution 

0.1

    Between -199.9 to +199.9

   

1

    Above 200

 (Auto Ranging) 

Accuracy(at 25

± 0.1% of rdg+0.2

  ± 0.1% of rdg+1

 ( Above 200

Sampling rate 

Approx. 0.5 sec/reading 

RS-232 output 

0.1

/1Bit (- 200.0 ~ +1800.0

) , Baud Rate=57600 

Power supply 

9V006P Battery × 1    About 160 hours for continuing use 
or DC-9V Adaptor (option) 

Dimension/Weight 

140 x 56 x 32 mm (L x W x D), Approx.120g (without battery) 

Notice: RS-232 transmission can not configure a variety of 

instruments on the same port.

 

K                                     

0000 

0100 

0001 

0101 

0010 

0110 

0011 

0111 

NORMAL 

00 

HOLD 

01 

MAX 

10 

MIN 

11 