APPENDIX
A-
IMR01N12-E7
28
Power
Power supply voltage:
90 to 264 V AC [Including power supply voltage variation]
(Rating 100 to 240 V AC), 50/60 Hz
24 V AC
10 % [Including power supply voltage variation]
(Rating 24 V AC), 50/60 Hz
24 V DC
10 % [Including power supply voltage variation]
(Rating 24 V DC)
Specify
when
ordering
Power consumption:
HA430: 16.5 VA max. (at 100 V AC)
22.5 VA max. (240 V AC)
15.0 VA max. (at 24 V AC)
430 mA max. (24 V DC)
HA930: 17.5 VA max. (at 100 V AC)
24.0 VA max. (240 V AC
16.0 VA max. (at 24 V AC)
470 mA max. (24 V DC)
General specifications
Insulation resistance:
Between measuring terminal and grounding:
20 M
or more at 500 V DC
Between power terminal and grounding:
20 M
or more at 500 V DC
Between measuring terminal and power terminal: 20 M
or more at 500 V DC
Withstand voltage:
Between measuring terminal and grounding:
1 minute at 1000 V AC
Between power terminal and grounding:
1 minute at 1500 V AC
Between power and measuring terminals:
1 minute at 2300 V AC
Power failure:
A power failure of 20 ms or less will not affect the control action.
If power failure of more than 20 ms occurs, controller will reset.
HOT start (1, 2) or COLD start (selectable).
Memory backup:
Backed up by non-volatile memory (EEPROM)
Number of writing:
Approx. 100,000 times
Data retaining period: Approx. 10 years
Depending on storage and operating conditions.
Allowable ambient temperature:
10 to
50
C (14 to 122
F)
Allowable ambient humidity:
5 to 95 %RH
Absolute humidity: MAX.W.C 29 g/m
3
dry air at 101.3 kPa
Ambient atmosphere:
There should be neither corrosive gases nor much dust.
Dimensions:
HA430: 48
(W)
96 (H)
100 (D) mm
HA930: 96
(W)
96 (H)
100 (D) mm
Weight:
HA430: Approx.
360
g
HA930:
Approx.
460
g
Standard
Safety standards:
UL:
UL61010-1
cUL:
CAN/CSA-C22.2 No.61010-1
CE marking:
LVD:
EN61010-1
OVERVOLTAGE
CATEGORYII,
POLLUTION
DEGREE
2,
Class II (Reinforced insulation)
EMC:
EN61326-1
RCM:
EN55011
Panel sealing:
IP65