SC-04
Chapter 4: Display Mode • 29
Changing the Station ID (STN ID)
This screen only appears if
DISP.SET
in User Mode is set to
ON
(factory setting) and if
ID
DISP
in Maintenance Mode is set to
ON
(factory setting is
OFF
).
Use this screen to select a station ID from the 128 station IDs that are stored in the SC-04’s
memory. Before a station ID is selected on a brand new instrument, the station ID is
“----------”. The factory-installed station IDs have a “S_ID_XXX” format.
The station ID provides a way to identify the SC-04 location during a data logging session.
User IDs can only be selected in this menu item. In order to edit the 128 user IDs, you must
use the 04 Series Datalogging Program or 04 Series Setup Program.
1. After entering Display Mode, press and release POWER MODE until the
STN ID
screen
sequence appears.
2. To change the Station ID, press and release AIR. The current Station ID flashes.
3. Use AIR to scroll to the desired Station ID.
4. Press and release POWER MODE to save the Station ID and return to the
STN ID
screen
in Display Mode.
Last Successful Calibration Date (CAL.DATA)
The
CAL.DATA
screen shows the date of the last successful calibration. This screen only
appears if
CAL.RMDR
is set to
ON
(factory setting).
1. After entering Display Mode, press and release POWER MODE until the
CAL.DATA
screen sequence appears.
STN
ID
S_ID_001
STN
ID
CHG / AIR
STN
ID
NO /MODE
STN
ID
S_ID_001
DISP
CAL .DATA
DISP
YES / AIR
DISP
NO /MODE