35-3001A-05-XX CO
2
Sample-Draw Detector • 23
7. Unscrew the regulator from the 100% nitrogen calibration cylinder.
Setting the Response Reading
1. Screw the regulator into the carbon dioxide calibration cylinder.
2. Connect the sample tubing from the demand flow regulator to the sample-draw detector’s inlet line.
3. Allow the sample-draw detector to draw sample for one minute.
4. When the reading on the voltmeter stabilizes, verify that the reading matches the response reading (±2
mV) you determined earlier.
5. If necessary, use the span potentiometer on the amplifier to adjust the reading to match the correct
response reading.
6. Disconnect the sample tubing from the inlet line.
7. Unscrew the regulator from the carbon dioxide calibration cylinder.
Returning to Normal Operation
1. Wait approximately one minute to allow the carbon dioxide reading to stabilize.
2. Remove the voltmeter leads from the amplifier test points.
3. Close the housing door.
4. Follow the instructions in the controller’s operator’s manual to exit the calibration mode.
5. Store the components of the calibration kit in a safe and convenient place.