35-3010A-13 Sample-Draw Detector • 11
Oxygen Transmitter
The oxygen transmitter is mounted to the right of the H
2
transmitter and above the main circuit board.
The amplifier includes the amplifier type selector, two internally wired terminal strips, span pot, zero
pot, and test points.
Figure 4: Oxgyen Transmitter Component Location
Transmitter Type Selector
The transmitter type selector is near the bottom left corner of the amplifier. It is to the left of the
detector terminal strip and below the span pot.
The transmitter included with the sample-draw detector is designed for use with RKI’s oxygen and
toxic gas sensors. The transmitter type selector determines for which sensor the amplifier is intended.
For oxygen transmitters, a jumper block is installed over the
OXY
selector.
Detector Terminal Strip
The detector terminal strip is the four-point terminal strip near the bottom edge of the transmitter. It is
factory wired to the transmitter.
Interconnect Terminal Strip
The interconnect terminal strip is the four-point terminal strip near the top edge of the transmitter. It is
factory wired to the main circuit board.
Span/Zero Pots
The span and zero pots are located on the left side of the transmitter and are used for calibration. Use
the span pot to make adjustments to gas response readings and the zero pot to make adjustments to the
zero reading
Test Points
The test points are located on the top of the transmitter on either side of the transmitter interconnect
terminal strip. They are labeled TP+ and TP-. A 100 mV - 500 mV output is available at these test
points for use during calibration.
TP
-
TP
+
4/20
FB
+
+
BATT
24V
Sensor Terminal
Strip
BK
W
OXY
SPAN
TOXIC
RD
TOXIC
OXY
G
Interconnect
Terminal Strip
Test Point TP-
Transmitter
Type Selector,
Set To OXY
Span Pot
ZERO
Test Point TP+
Zero Pot