14 • 35-3010A-13 Sample-Draw Detector
Installation
This section describes procedures to mount the sample-draw gas detector in the monitoring
environment and wire the sample-draw detector to power and an external device.
Mounting the Sample-Draw Detector
1. Select the mounting site. Consider the following when you select the mounting site.
•
Is there enough room to open the housing door and make wiring connections at the bottom of
the housing and tubing connections at the right of the housing?
•
Make sure there is sufficient room to perform start-up, maintenance, and calibration
procedures.
Figure 6: Outline and Mounting Dimensions
2. Close and latch the housing door.
NOTE:
The sample-draw detector is shipped with the mounting feet “tucked under” the housing to
protect the mounting feet during shipment.
Ø .31 x .50 slot, 4X
8.31
.18
Tube Fitting for 1/4" OD
Tube, 2X
3/4" Conduit Hub, 2X
12.55
Door Latch
10.00
14.94
14.55
15.44