EAGLE 2 Operator’s Manual
Using Setup Mode • 105
14. Press and release POWER ENTER RESET. The confirmation screen appears. In the
example below, the user defined gas has been selected and defined as propane with
the gas name set to PRO.
15. If you want to accept the gas configuration change, press and release AIR
▲
YES.
The unit will return to the Configure Gases screen.
If you do not want to accept the gas configuration change, press and release
DISPLAY ADJUST NO to return to the screen with the gas choices shown in Step 4
on page 101. You can either scroll down to
END
and press POWER ENTER RESET
to return to the Configure Gases screen or continue from Step 4 on page 101 to select
a new gas.
16. Use RANGE
▼
SHIFT to place the cursor next to
END
and press POWER ENTER
RESET to return to the main menu.
17. Be sure to perform a calibration on the catalytic combustible channel using an
appropriate calibration cylinder for the target gas.
Setting the Catalytic Detection Units
This menu item allows you to display the combustible gas units on the catalytic sensor
channel as ppm, %LEL, %vol, or selectable between the three units (the
CHANGE OK
option). The factory setting is
CHANGE OK
.
1. From the main menu, place the cursor next to
CATALYTIC UNITS
.
2. Press and release POWER ENTER RESET. The Catalytic Units screen appears with
the current setting flashing at the bottom of the screen.
3. Use AIR
▲
YES or RANGE
▼
SHIFT to scroll through the choices,
CHANGE OK
,
vol% ONLY
,
%LEL ONLY
, and
ppm ONLY
.
4. When the desired setting is on the screen, press and release POWER ENTER RESET
to save the setting and return to the main menu.
CHANGE TO PRO ?
PRESS YES OR NO
CONFIGURE GASES
> CAT : PRO (CAT)
OP1 : --- (---)
OP2 : --- (---)
OP3 : --- (---)
CATALYTIC UNITS
UP/DOWN THEN ENTER
CHANGE OK