EAGLE 2 Operator’s Manual
Using Setup Mode • 107
4. Move the cursor next to the alarm point or alarm operation (oxygen only) that you
want to update.
If you selected the oxygen channel, you can set the alarm rising/falling operation in
addition to the alarm points.
NOTE:
If Inert Mode is active, you can change the oxygen alarm points for both
Normal Mode and Inert Mode. For more information about changing Inert
Mode alarm settings, see “Appendix N: Using the EAGLE 2 in Inert Mode” on
5. Press and release POWER ENTER RESET. The alarm point or alarm operation
(oxygen only) will begin to flash.
6. Use AIR
▲
YES and RANGE
▼
SHIFT to adjust the alarm point or alarm operation
(oxygen only) to the desired setting. Keep the following in mind:
•
The low alarm cannot be set higher than the high alarm and the high alarm cannot
be set lower than the low alarm.
•
Any alarm setting can be turned off by adjusting it to its lowest setting. The
setting will be displayed as
OFF.
•
In addition to setting the oxygen alarm points, you can also select one of the
following operation modes: low alarm decreasing and high alarm increasing
(
FALLING AND RISING
); low and high alarm decreasing (
BOTH
FALLING
); low and high alarm increasing (
BOTH RISING
). The factory
setting is
FALLING AND RISING
.
•
In order to appropriately set the calibration gas value during a calibration, the low
alarm setting must be lower than the desired calibration value.
7. If you want to continue with the change, press and release POWER ENTER RESET
to accept the setting.
If you want to exit this screen without saving any change to the alarms, press and
release DISPLAY ADJUST NO until you return to the Change Alarm Point Settings
Screen.
8. Repeat Step 4- Step 7 for any additional changes you want to make.
9. When you are done making changes, use RANGE
▼
SHIFT to move the cursor next
to
END
.
10. Press and release POWER ENTER RESET to save the new settings and return to the
Change Alarm Point Settings Screen.
11. Use RANGE
▼
SHIFT to move the cursor next to
END.
12. Press and release POWER ENTER RESET to return to the main menu.
OXY 0- 40.0 vol%
>FALLING AND RISING
LO ALRM : 19.5 vol%
HI ALRM : 23.5 vol%
END