108 • Using Setup Mode
EAGLE 2 Operator’s Manual
Updating the Alarm Latching Setting
With
ALARM LATCHING
set to
LATCHING
(factory setting), the EAGLE 2 remains
in alarm condition until the alarm condition passes
and
the POWER ENTER RESET
button is pressed.
With
ALARM LATCHING
set to
SELF RESET
, the EAGLE 2 automatically resets an
alarm when the alarm condition passes.
1. From the main menu, place the cursor next to
ALARM LATCHING
.
2. Press and release POWER ENTER RESET. The Alarm Latching Screen appears.
3. Use AIR
▲
YES or RANGE
▼
SHIFT to display the desired setting.
4. Press and release POWER ENTER RESET to save the setting and return to the main
menu.
Updating the Alarm Silence Setting
With
ALARM SILENCE
set to
ON
(factory setting), pressing and releasing any button
silences the buzzer when the EAGLE 2 is in alarm. The LEDs continue to flash and the
display continues to show the alarm. When the gas concentration falls below the alarm
level, pressing and releasing POWER ENTER RESET clears all alarm indications for that
alarm. If an alarm condition occurs, you may still enter Display Mode by pressing the
DISPLAY button. The buzzer will be silenced but the LEDs will continue to flash. If you
return to Measuring Mode and there is still an alarm condition, the LEDs will continue to
flash and the buzzer will remain off. Once the condition clears, press POWER ENTER
RESET to clear the alarm indications.
With
ALARM SILENCE
set to
OFF
, you cannot silence the buzzer. If an alarm
condition occurs, and you enter Display Mode, the buzzer will not be silenced and the
LEDs will continue to flash. Upon return to Measuring Mode, if there is still an alarm
condition, you must wait until it clears before you can press POWER ENTER RESET to
clear the alarm indications.
1. From the main menu, place the cursor next to
ALARM SILENCE
.
2. Press and release POWER ENTER RESET. The Alarm Silence Option Screen
appears.
3. Use AIR
▲
YES or RANGE
▼
SHIFT to display the desired setting.
4. Press and release POWER ENTER RESET to save the setting and return to the main
menu.
ALARM LATCHING
UP/DOWN THEN ENTER
LATCHING
ALARM
SILENCE OPTION
UP/DOWN THEN ENTER
ON
www.
.com
information@itm.com
1.800.561.8187