110 • Using Setup Mode
EAGLE 2 Operator’s Manual
5. Use AIR
▲
YES and RANGE
▼
SHIFT to adjust the auto calibration value to the
desired value.
6. Press and release the POWER ENTER RESET button to accept the value.
7. Repeat Step 3 - Step 6 for each auto calibration value you want to change. If you want
to return to the main menu at any time without saving any changes, press and release
the DISPLAY ADJUST NO button until you return to the main menu.
8. Use RANGE
▼
SHIFT to move the cursor next to the
END
menu item. Press and
release the POWER ENTER RESET button to save the changes and return to the
main menu.
Updating the Backlight Delay Setting
This setting indicates the length of time the LCD illuminates when you press any button.
The minimum setting is 0 seconds; the maximum setting is 255 seconds. The factory
setting is 30 seconds.
1. From the main menu, place the cursor next to
BACKLIGHT DELAY
.
2. Press and release POWER ENTER RESET. The Backlight Delay Screen appears.
3. Use AIR
▲
YES and RANGE
▼
SHIFT to adjust the time to the desired setting.
4. Press and release POWER ENTER RESET to save the setting and return to the main
menu.
Updating the Auto Fresh Air Setting
This setting allows you to configure the EAGLE 2 so that a fresh air adjustment takes
place automatically as part of the instrument startup sequence. If
AUTO FRESH AIR
ADJ
is set to
ON
, the EAGLE 2 performs a fresh air adjustment at the end of the startup
sequence before entering Normal Operation. The factory setting is
OFF
.
WARNING:
If the automatic fresh air feature is turned on, you must startup the
EAGLE 2 in a known fresh air environment, an environment free of
toxic or combustible gases and of normal oxygen content (20.9%). If
this feature is on and the EAGLE 2 is started up in the presence of a
target gas, the readings and alarms will not be accurate or reliable.
1. From the main menu, place the cursor next to
AUTO FRESH AIR ADJ
.
2. Press and release POWER ENTER RESET. The Fresh Air Adjust Screen appears.
BACKLIGHT DELAY
UP/DOWN THEN ENTER
30 SECONDS
FRESH AIR ADJUST
ON POWERUP
UP/DOWN THEN ENTER
OFF
www.
.com
information@itm.com
1.800.561.8187