EAGLE 2 Operator’s Manual
Using Setup Mode • 113
Updating the Calibration Reminder Setting
With
CAL REMINDER
set to
ON
(factory setting), the EAGLE 2 will give an indication
at start up if it is due for calibration. The type of indication will depend on the
CAL PAST
DUE ACT
setting (see the next menu item below).
1. From the main menu, place the cursor next to
CAL REMINDER
.
2. Press and release POWER ENTER RESET. The Calibration Reminder Screen
appears.
3. Use AIR
▲
YES or RANGE
▼
SHIFT to display the desired setting.
4. Press and release POWER ENTER RESET to save the setting and return to the main
menu.
Updating the Calibration Past Due Action Setting
This item defines what indication is given during start up when calibration is due and
CAL REMINDER
is set to
ON
.
With
CAL PAST DUE ACT
set to
CONFIRM TO CAL
(factory setting), the EAGLE 2
will give an indication at start up if calibration is past due and require the user to decide
whether to perform a calibration or continue and use the EAGLE 2 without calibrating.
Press and release DISPLAY ADJUST NO to continue without calibrating or AIR
▲
YES
to perform a calibration.
With
CAL PAST DUE ACT
set to
MUST CALIBRATE
, if the unit is due for
calibration, the EAGLE 2 will give an indication at start up that calibration is past due and
prompt you to press and release POWER ENTER RESET to enter Calibration Mode and
perform a calibration. Using any other button will have no effect. A successful calibration
must be performed in order to use the instrument.
NOTE:
Even if the password function is turned on in
CHANGE PASSWORD
, no
password will be required to perform a calibration during startup.
With
CAL PAST DUE ACT
set to
NOTIFICATION ONLY,
the EAGLE 2 will give an
indication at startup that calibration is past due. You must press and release POWER
ENTER RESET to acknowledge the indication and proceed with the startup sequence.
1. From the main menu, place the cursor next to
CAL PAST DUE ACT
.
2. Press and release POWER ENTER RESET. The Calibration Past Due Action Screen
appears.
CALIBRATION
REMINDER
UP/DOWN THEN ENTER
ON
CALIBRATION PAST
DUE ACTION
UP/DOWN THEN ENTER
CONFIRM TO CAL