124 • Channel Setup & Sub PCBs
EAGLE 2 Operator’s Manual
•
The sensor type
The sensor type can be one of the following: catalytic (CAT), photo ionization
detector (PID), ESM-01 toxic (ESM), infrared (IR), or thermal conductivity
(TC). Although any installed sensor type will be listed in CONFIGURE GASES,
the only sensor types whose gas can be configured are catalytic, PID, and TC
sensor types.
If you select an ESM-01 toxic or IR type of sensor to configure in this menu
item, a message will appear indicating that the gas configuration for this sensor
is automatically detected by the EAGLE 2 and that manual configuration is not
possible. Both of these sensor types have the gas configuration stored in memory
onboard the sensor.
See the appropriate optional sensor appendix for a complete description of configuring the
gas for that type of sensor.
ESM & IR SENSOR
ARE AUTOMATICALLY
DETECTED.
MANUAL SELECTION