EAGLE 2 Operator’s Manual
Maintenance • 151
Steps to Replace the Lamp
Perform the following procedure to replace the PID lamp:
NOTE:
Do not touch the new lamp window (the flat end) with your fingers as this may
contaminate the window with finger oil.
1. Verify that the EAGLE 2 is off.
2. Place the EAGLE 2 upside down on a flat surface or hold it upside down.
3. Unscrew the three case screws until they disengage from the top case. They are
captive screws so they will not fall off of the bottom case.
4. Turn the EAGLE 2 right side up and carefully lift the top case away from the bottom
case. Be careful not to lift it so far that it pulls on the main PCB with the cable that
connects the top case to the main PCB.
5. Lay the top case down next to the bottom case to allow access to the flow system.
6. Locate the PID sensor adapter. It has a five wire cable and is normally located next to
the pump. The cable has a connector that mates to a PID sub PCB that is installed on
the main PCB. Figure 27 on page 126 shows a PID sensor in a typical location.
7. Grasp the sensor adapter firmly and pull it out of the PID flow chamber. Rock it back
and forth gently if necessary to pull it out. Take care not to pull the cable connector
from the PID sub PCB.
8. The PID sensor protrudes from one end of the sensor adapter. Grasp the PID sensor
firmly and pull it out of the sensor adapter.
9. Place the PID sensor face down on a flat clean working surface.
10. Hold the PID sensor steady on the working surface with one hand and using the other
hand, locate the tabs on the electrode stack removal tool and insert them into the slots
on the side of the PID sensor near the face.
Figure 33: Using Removal Tool