154 • Maintenance
EAGLE 2 Operator’s Manual
10. Hold the PID sensor steady on the working surface with one hand and using the other
hand, locate the tabs on the electrode stack removal tool and insert them into the slots
on the side of the PID sensor near the face.
Figure 35: Using Removal Tool
11. Squeeze the removal tool to push the tabs into the sensor slots until the electrode stack
and lamp are released.
12. Carefully lift the PID sensor body away from the electrode stack and lamp. If the
lamp remains lodged in the sensor body, carefully remove it with tweezers.
13. If the spring in the lamp cavity comes out, place it back into the lamp cavity.
14. Discard the old electrode stack.
15. Hold the new electrode stack between the thumb and forefinger of one hand and place
the window end of the lamp inside the O-ring seal in the new electrode stack with the
other hand as shown below. Take care not to touch the lamp window.
www.
.com
information@itm.com