EAGLE 2 Operator’s Manual
ESM-01 Calibration • 175
8. Press and release the AIR
▲
YES button to continue.
If you do not want to continue, press the DISPLAY ADJUST NO button and the unit
will return to the Calibration Mode Screen.
9. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds,
then indicate that the operation is complete before returning to the Calibration Mode
Screen.
10. Install the demand flow regulator onto the multi-gas calibration cylinder.
11. Connect the sample tubing to the demand flow regulator.
12. Install the probe on the EAGLE 2 inlet fitting. Make sure the probe is complete with
internal O-ring and membrane and that the two halves of the probe are tightened
firmly together to avoid leaks that can affect the calibration. See Figure 21 on page 72
for an illustration of the internal parts of the probe.
13. Move the cursor next to the
AUTO CALIBRATION
menu item by using the AIR
▲
YES button.
ADJUSTING ZERO
RELEASE AIR BUTTON
ADJUSTING ZERO
COMPLETE
CALIBRATION MODE
AUTO CALIBRATION
SINGLE CALIBRATION
> PERFORM AIR ADJUST
NORMAL OPERATION
CALIBRATION MODE
> AUTO CALIBRATION
SINGLE CALIBRATION
PERFORM AIR ADJUST
NORMAL OPERATION