EAGLE 2 Operator’s Manual
Configuring the TC Gas in Setup Mode • 195
Configuring the TC Gas in Setup Mode
The TC channel can be configured for 3 different pre-defined gases and 1 user-defined gas
in the CONFIGURE GASES menu item in Setup Mode. To change the gas configuration
of the TC channel in Setup Mode, do the following:
WARNING:
Do not use the user defined gas configuration without consulting RKI
Instruments, Inc.
WARNING:
The EAGLE 2 is not in operation as a gas detector while in Setup Mode.
1. Take the EAGLE 2 to a non-hazardous location and turn it off if it is on.
2. Press and hold AIR
▲
YES and RANGE
▼
SHIFT, then press and hold POWER
ENTER RESET. When you hear a beep, release the buttons.
3. The LCD will show the following screen for a few seconds with the “S” in the lower
right corner indicating the unit is entering Setup Mode.
4. The “S” will then disappear and the following screen will appear for a few seconds.
5. If the unit prompts you for the password, enter it by using AIR
▲
YES and RANGE
▼
SHIFT to select each password number and then pressing and releasing the
POWER ENTER RESET button to enter it and move on to the next number until all
of the numbers are entered. The main menu displays. It displays six menu items at a
time.
6. Use the RANGE
▼
SHIFT button to move the cursor down the menu to
CONFIGURE GASES.
GAS MONITOR
EAGLE 2 S
GAS MONITOR
EAGLE 2
>SET DATE & TIME
SET DATE FORMAT
SET BATTERY TYPE
CONFIGURE CHANNELS
CONFIGURE GASES
CATALYTIC UNITS
www.
.com
information@itm.com
1.800.561.8187