EAGLE 2 Operator’s Manual
Configuring the TC Gas in Setup Mode • 197
13. The user defined gas setup screen appears with the first character of the gas name
flashing. The current gas name and range are shown on the top line of the screen.
WARNING:
Do not use the user defined gas configuration without consulting RKI
Instruments, Inc.
14. Enter the gas name. There are 3 characters available for the gas name. The factory
setting of TC uses only 2 characters. Use AIR
▲
YES and RANGE
▼
SHIFT to
display the desired character, then press POWER ENTER RESET to enter the
displayed character and move to the next character. Repeat until all three characters
are entered. When the last character is entered, the full scale value will be flashing.
15. Use AIR
▲
YES and RANGE
▼
SHIFT to display the desired full scale value. It may
be any value from 10 to 100 in increments of 10.
16. Press and release POWER ENTER RESET to enter the full scale value. The
confirmation screen appears. In the example below, the user defined gas has been
selected and defined as propane with the gas name set to PRO.
17. If you want to accept the gas configuration change, press and release AIR
▲
YES.
The unit will return to the Configure Gases screen.
If you do not want to accept the gas configuration change, press and release
DISPLAY ADJUST NO to return to the screen with the gas choices shown in Step 10
on page 196. You can either scroll down to END and press POWER ENTER RESET
to return to the Configure Gases screen or continue from Step 10 on page 196 to
select a new gas.
18. Use RANGE
▼
SHIFT to place the cursor next to
END
.
19. Press and release POWER ENTER RESET to return to the main menu.
TC (0-100.0vol%CH4)
CHANGE TO NAME
TC
FULL_SCALE 10.0
CHANGE TO PRO ?
PRESS YES OR NO
CONFIGURE GASES
> CAT : CH4 (CAT)
OP1 : PRO (TC)
OP2 : --- (---)
OP3 : --- (---)