22 • Standard Accessories
EAGLE 2 Operator’s Manual
are available from 5 feet (standard length) to 125 feet (see “General Parts List” on
page 79). A Teflon lined hose is provided with all units that contain a PID sensor. This
hose must be used when operating a PID EAGLE 2 (see “Appendix D: PID Sensors” on
CAUTION:
Sample hose lengths of more than 125 feet are not recommended for the
EAGLE 2 because of flow rate reduction and increased response time.
Consult RKI Instruments, Inc. for hose lengths longer than 125 feet.
The probe includes a replaceable particle filter and hydrophobic filter disk that prevent
particulates and water from entering the EAGLE 2’s flow system. See “Replacing the
Hydrophobic Probe’s Particle Filter and Hydrophobic Filter Disk” on page 72 for
instructions to replace the particle filter and hydrophobic filter disk.
NOTE:
When using the probe with a PID EAGLE 2, be sure that the particle filter is not
installed.
Figure 6: Sample Hose and Hydrophobic Probe
Sample Hose
Probe
www.
.com
information@itm.com
1.800.561.8187