232 • Infrared Hydrocarbon Calibration
EAGLE 2 Operator’s Manual
21. When you are done adjusting the calibration gas values, move the cursor down past
the bottom of the screen next to
END
.
22. Press and release the POWER ENTER RESET button. The following screen appears.
23. If you select YES by pressing and releasing the AIR
▲
YES button, the changes that
you made will be saved in the EAGLE 2’s memory as the new auto calibration gas
values.
If you select NO by pressing and releasing the DISPLAY ADJUST NO button, the
changes you made will be used for any calibrations performed during the current
operating session only. The EAGLE 2 will delete the changes when the unit is turned
off and will load the previous set of auto calibration values when it is turned on again.
24. When you make your selection and press the desired button, the unit returns to the
Calibration Gas Values Screen.
25. Press and release the POWER ENTER RESET button to proceed to the Calibration In
Process Screen with
CAL IN PROCESS
flashing.
If you do not want to proceed with the calibration, press and release the DISPLAY
ADJUST NO button to return to the Cal Gas Values Screen.
If you do want to continue with the calibration, proceed to the next step.
ADJUST AUTO
CALIBRATION VALUES
> END
DO YOU WANT TO
STORE NEW VALUE(S)
IN MEMORY FOR
FUTURE CALIBRATIONS?
PRESS YES OR NO
CAL GAS VALUES
CH4 50 %LEL
OXY 12.0 vol%
H2S 25.0 ppm
CO 50 ppm
ENTER TO BEGIN CAL
CAL IN PROCESS
CH4 0 %LEL
OXY 20.9 vol%
H2S 0.0 ppm
CO 0 ppm
ENTER WHEN DONE