EAGLE 2 Operator’s Manual
Overview • 241
Appendix K: Using the EAGLE 2 in Bar Hole Mode
Overview
This chapter explains how to operate the EAGLE 2 in Bar Hole Mode. Bar Hole Mode is
used to perform consistent checks of bar holes when tracking down underground gas
leaks. When the EAGLE 2 is in Bar Hole Mode, only the combustible and oxygen sensors
are displayed.
If an EAGLE 2 is intended for bar hole testing, it is shipped with Bar Hole Mode or both
Bar Hole Mode and Leak Check Mode enabled so that the operator must choose which
operational mode to use when the unit is turned on (see “Updating LC/BH Mode Setting”
Start Up, Bar Hole Mode
This section explains how to start up the EAGLE 2 in Bar Hole Mode and get it ready for
operation.
Turning On the EAGLE 2, Bar Hole Mode
WARNING:
If one or more sensors other than a catalytic LEL, TC, or oxygen
sensor is installed, these sensors will not be active while the
EAGLE 2 is in Bar Hole Mode.
WARNING:
Gas alarms are not active when the EAGLE 2 is in Bar Hole Mode.
The following description of the EAGLE 2 start up sequence assumes that the following
menu items in Setup Mode are turned on: CAL REMINDER and USER/STATION ID. If
either of these items is turned off, then the corresponding screen will not appear.
NOTE:
In order for
BAR HOLE MODE
to appear as a selection in the Mode Select
Screen in Step 3 below, the EAGLE 2 must have both a catalytic LEL sensor
and a TC sensor installed. In addition, both the catalytic LEL and TC channels
must be configured for methane, CH
4
, in the CONFIGURE GASES menu item
in Setup Mode. If an EAGLE 2 has only the catalytic LEL sensor installed or if
the TC channel is configured for a gas other than CH
4
, then
BAR HOLE
MODE
will not appear as a choice as shown in Step 3 below and
NORMAL
MODE
and
LEAK CHECK MODE
will be the only choices displayed. In this
case, see “Turning On the EAGLE 2” on page 27 or “Turning On the EAGLE 2,
1. Connect the sample hose to the EAGLE 2’s quick connect inlet fitting.
2. Connect the bar hole probe to the sample hose’s quick connect fitting.
3. Press and briefly hold down the POWER ENTER RESET button. Release the button
when you hear a beep.