244 • Start Up, Bar Hole Mode
EAGLE 2 Operator’s Manual
•
If the unit is due for calibration and CAL PAST DUE ACT is set to MUST
CALIBRATE, then the following screen displays and the buzzer sounds in a
double pulsing pattern.
The EAGLE 2 cannot be used until a successful calibration has been performed.
Press and release the ENTER button to enter Calibration Mode. See “Chapter 4:
Calibration Mode” on page 54 for instructions to calibrate the EAGLE 2. When
you are done with the calibration and exit Calibration Mode, the unit will begin
the startup sequence. If the calibration was successful, the screen above will not
appear again until the unit is due for calibration. If the calibration was not
successful, the screen above will again appear in the startup sequence.
•
If the unit is due for calibration and CAL PAST DUE ACT is set to
NOTIFICATION ONLY, then the following alert screen displays and the buzzer
sounds in a double pulsing pattern.
Press and release the POWER ENTER RESET button to acknowledge the alert
and continue with the startup sequence.
12. The Date/Time Screen appears for a few seconds.
13. If USER/STATION ID is turned on (see “Turning the User/Station ID Function On or
Off” on page 109), the ID Screen appears for a few seconds.
If USER/STATION ID is turned off, only the serial number is shown.
CALIBRATION DATE
IS PAST DUE
ENTER TO PERFORM
CALIBRATION
CALIBRATION DATE
IS PAST DUE
9/12/2008
15:00:00
USER ID
MIKE
STATION ID
PUMP 1
SERIAL NUMBER
E2A515