260 • Calibration
EAGLE 2 Operator’s Manual
Calibration
Use a hexane calibrating source to calibrate the combustible gas LEL range. Use a 100%
nitrogen calibrating source to set the zero reading for the oxygen channel. RKI
Instruments, Inc. recommends using the Single Calibration method to calibrate the
EAGLE 2 Tank Tester model. See “Chapter 4: Calibration Mode” on page 54. The
instructions in that section call for calibrating the EAGLE 2 with methane. When
following these instructions, be sure to calibrate with hexane rather than methane.
NOTE:
Do not calibrate the EAGLE 2 Tank Tester model with the dilution fitting
attached to the inlet fitting.
Parts List
Table 36 lists part numbers for replacement parts and accessories of the EAGLE 2’s Tank
Tester model.
Table 36: Parts List: EAGLE 2 Tank Tester Model
Part Number
Description
80-0405RK
Dilution fitting (1:1)
80-0802RK
Float probe (12-foot)