28 • Start Up
EAGLE 2 Operator’s Manual
5. The Battery Voltage Screen appears for a few seconds.
6. The Active Gases Screen appears for a few seconds indicating which channels are
active and their target gas.
7. If LUNCH BREAK is turned on (see “Updating the Lunch Break Setting” on
page 118), the Resume Measurement Screen appears. The unit counts down from 5
seconds in the lower right corner of the LCD to the right of “MEASUREMENTS”.
•
To continue accumulating peak and time-weighted average (TWA) readings from
the last time the EAGLE 2 was used, press and release the AIR
▲
YES button
before the countdown reaches 0 or allow the countdown to reach 0. If you do not
press the AIR
▲
YES button within the 5 second countdown, the EAGLE 2
automatically resumes accumulating the peak and TWA readings.The EAGLE 2
will also continue to keep track of operating time including the operating time
from the last time the EAGLE 2 was used. See “Time in Operation Screen” on
page 50 for more information about how the EAGLE 2 tracks the operating time.
The short-term exposure limit (STEL) reading is reset each time the EAGLE 2 is
turned on.
•
To reset the accumulation of these measurements, press and release the
DISPLAY ADJUST NO button before the countdown reaches 0.
BATTERY MIN:
4.3 VOLTS
BATTERY NOW:
5.2 VOLTS
ACTIVE GASES
CH4 OXY
H2S CO
LUNCH BREAK MODE ON
RESUME
PEAK AND TWA
MEASUREMENTS? 2