EAGLE 2 Operator’s Manual
Start Up • 29
8. The gas alarm setpoints are displayed by three screens in sequence: the Low Alarm
Screen, High Alarm Screen, and STEL/TWA Alarm Screen. Each screen remains on
the LCD for three seconds.
9. After the alarm screens, if CAL REMINDER is turned on, the screen that appears
next depends on how CAL PAST DUE ACT is set in the Setup Mode Menu (see
“Updating the Calibration Past Due Action Setting” on page 113).
•
If the unit is due for calibration and CAL PAST DUE ACT is set to CONFIRM
TO CAL, then the following screen displays and the buzzer sounds in a double
pulsing pattern.
To perform a calibration, press and release the AIR
▲
YES button. The EAGLE
2 will enter Calibration Mode and the LCD will show the Calibration Mode main
menu. See “Chapter 4: Calibration Mode” on page 54 for instructions to calibrate
the EAGLE 2. When you are done with the calibration and exit Calibration
Mode, the unit will begin the startup sequence. If the calibration was successful,
the screen above will not appear again until the unit is due for calibration. If the
calibration was not successful, the screen above will again appear in the startup
sequence.
To continue without performing a calibration, press and release the DISPLAY
ADJUST NO button.
A CH4 10 %LEL
L L OXY 19.5 vol%
O A H2S 10.0 ppm
WR CO 25 ppm
M
S
CH4 10 vol%
A CH4 50 %LEL
HL OXY 23.5 vol%
I A H2S 30.0 ppm
GR CO 50 ppm
HM
S
CH4 50 vol%
ALARMS STEL & TWA
H2S(ppm) 15.0 10.0
CO (ppm) 200 25
CALIBRATION DATE
IS PAST DUE
PERFORM
CALIBRATION?