30 • Start Up
EAGLE 2 Operator’s Manual
•
If the unit is due for calibration and CAL PAST DUE ACT is set to MUST
CALIBRATE, then the following screen displays and the buzzer sounds in a
double pulsing pattern.
The EAGLE 2 cannot be used until a successful calibration has been performed.
Press and release the ENTER button to enter Calibration Mode. See “Chapter 4:
Calibration Mode” on page 54 for instructions to calibrate the EAGLE 2.
NOTE:
In this situation, even if the password function has been turned on, no password
is required to perform a calibration.
When you are done with the calibration and exit Calibration Mode, the unit will
begin the startup sequence. If the calibration was successful, the screen above
will not appear again until the unit is due for calibration. If the calibration was
not successful, the screen above will again appear in the startup sequence.
•
If the unit is due for calibration and CAL PAST DUE ACT is set to
NOTIFICATION ONLY, then the following alert screen displays and the buzzer
sounds in a double pulsing pattern.
Press and release the POWER ENTER RESET button to acknowledge the alert
and continue with the startup sequence.
10. The Date/Time Screen appears for a few seconds.
CALIBRATION DATE
IS PAST DUE
ENTER TO PERFORM
CALIBRATION
CALIBRATION DATE
IS PAST DUE
9/12/2008
15:00:00