42 • Display Mode
EAGLE 2 Operator’s Manual
•
display alarm settings
•
select the user ID (if the
USER/STATION ID
item in Setup Mode is set to
ON
)
•
select the station ID (if the
USER/STATION ID
item in Setup Mode is set to
ON
)
•
display time in operation
•
display date and time
•
display remaining data logging time and clear data logger memory (if the
DATA LOG
MEMORY
item in Setup Mode is set to
ON
)
Tips for Using Display Mode
•
To enter Display Mode and scroll from one screen to the next or skip an item when a
question is asked, press and release the DISPLAY ADJUST NO button.
•
To enter an item when a question is asked, press and release the AIR
▲
YES button.
•
To change a flashing parameter, use either the AIR
▲
YES button or RANGE
▼
SHIFT button.
NOTE:
Each screen displays for 20 seconds. If you do not press a button within 20
seconds, the EAGLE 2 automatically returns to Measuring Mode.
Peak Screen
The peak screen displays the highest (lowest for oxygen) concentrations detected since the
EAGLE 2 was turned on. Peak readings are stored in the EAGLE 2’s memory until a
higher level is detected (lower for oxygen), the peak reading is cleared, or the EAGLE 2 is
turned off.
The lunch break feature enables the EAGLE 2 to save peak readings when it is turned off
so it can continue them when it is turned on again. See “Turning On the EAGLE 2” on
To clear the peak readings, do the following:
1. With the Peak Screen displayed, press and release the POWER ENTER RESET
button. The following screen will appear.
2. Press and release the AIR
▲
YES button. The peak readings will be reset and the unit
will return to the Peak Screen.
CH4 0%LEL
P OXY 20.9vol%
E H2S 0.0ppm
A CO 0ppm
K
CLEAR
PEAK READINGS?