EAGLE 2 Operator’s Manual
Display Mode • 43
If you do not want to clear the peak readings, press and release the DISPLAY
ADJUST NO button and the unit will return to the Peak Screen without clearing the
peak readings.
Battery Voltage Screen
The Battery Voltage Screen displays the minimum operating voltage and the current
battery voltage. Fully charged alkaline batteries typically indicate 6.0 volts; fully charged
Ni-MH batteries typically indicate 5.2 volts. This screen also displays during the startup
sequence.
Gas Display Screen
The Gas Display Screen gives you the option to select how the active channels are
displayed.
You can display all of them on the screen at the same time, one at a time with automatic
scrolling, or one at time with manual scrolling. The factory setting displays all of the
active channels at the same time. To select how to display the active channels, do the
following:
1. With the Gas Display Screen displayed, press and release the
AIR
▲
YES button. The following screen will appear with the cursor blinking.
2. Use the AIR
▲
YES and RANGE
▼
SHIFT buttons to place the cursor next to the
desired choice.
3. Press and release the POWER ENTER RESET button. The unit continues to the
STEL Screen and the display configuration will reflect your choice when you return
to Measuring Mode.
If you do not want to change the setting, press and release the DISPLAY ADJUST
NO button and the unit will return to the Gas Display Screen.
BATTERY MIN:
4.3 VOLTS
BATTERY NOW:
5.2 VOLTS
SELECT
GASES DISPLAYED
SELECT
> DISPLAY ALL
SCROLL AUTO
SCROLL MANUAL
www.
.com
information@itm.com
1.800.561.8187