EAGLE 2 Operator’s Manual
Calibrating Using the Auto Calibration Method • 61
19. If all channels passed calibration, the following screen sequence occurs.
If any of the sensors cannot be adjusted to the proper value, a screen displays that
indicates a calibration failure and lists the sensor(s) that failed to calibrate. In the
example below, the oxygen and H
2
S channels failed calibration. The other sensors
calibrated normally.
The buzzer and alarm LED arrays activate in a double pulsing pattern. Press and
release the POWER ENTER RESET button to reset the alarm and return to the
Calibration Mode Screen. Attempt to calibrate again. If the failure continues,
investigate the cause. See “Troubleshooting” on page 67.
20. Disconnect the tubing from the probe.
21. Unscrew the demand flow regulator from the calibration cylinder.
22. Use the RANGE
▼
SHIFT button to place the cursor next to the
NORMAL
OPERATION
menu option, then press and release the POWER ENTER RESET
button to return to Measuring Mode.
AUTO CALIBRATION
DONE
REMOVE GAS
AUTO CALIBRATION
END
CALIBRATION MODE
> AUTO CALIBRATION
SINGLE CALIBRATION
PERFORM AIR ADJUST
NORMAL OPERATION
FAIL SENSOR
< > <OXY>
<H2S> < >
ENTER TO CONTINUE