62 • Calibrating Using the Single Calibration Method
EAGLE 2 Operator’s Manual
Calibrating Using the Single Calibration Method
Single Calibration allows you to calibrate one channel at a time. This is useful if you only
want to calibrate one or two channels.
Setting the Fresh Air Reading
1. While in the Calibration Mode Screen, move the cursor to the
PERFORM AIR
ADJUST
menu item by using the RANGE
▼
SHIFT button.
2. Press and release the POWER ENTER RESET button. The following screen appears.
3. Press and release the AIR
▲
YES button to continue.
If you do not want to continue, press the DISPLAY ADJUST NO button and the unit
will return to the Calibration Mode Screen.
CALIBRATION MODE
AUTO CALIBRATION
SINGLE CALIBRATION
> PERFORM AIR ADJUST
NORMAL OPERATION
PERFORM
AIR ADJUST?