EAGLE 2 Operator’s Manual
Calibrating Using the Single Calibration Method • 63
4. The EAGLE 2 will indicate that it is adjusting the zero reading for a few seconds,
then indicate that the operation is complete before returning to the Calibration Mode
Screen.
Performing a Span Adjustment in Single Calibration
1. Install the demand flow regulator onto the calibration cylinder.
2. Connect the sample tubing to the demand flow regulator.
3. Install the probe on the EAGLE 2 inlet fitting. Make sure the probe is complete with
internal O-ring and membrane and that the two halves of the probe are tightened
firmly together to avoid leaks that can affect the calibration. See Figure 21,
“Replacing the Particle Filter and Hydrophobic Filter Disk” on page 72 for an
illustration of the internal parts of the probe.
4. Move the cursor next to the
SINGLE CALIBRATION
menu item by using the AIR
▲
YES button.
ADJUSTING ZERO
RELEASE AIR BUTTON
ADJUSTING ZERO
COMPLETE
CALIBRATION MODE
AUTO CALIBRATION
SINGLE CALIBRATION
> PERFORM AIR ADJUST
NORMAL OPERATION
CALIBRATION MODE
AUTO CALIBRATION
> SINGLE CALIBRATION
PERFORM AIR ADJUST
NORMAL OPERATION