66 • Calibrating Using the Single Calibration Method
EAGLE 2 Operator’s Manual
NOTE:
The adjustment range will only appear in the calibration results screen if the
Span Factor item in Setup Mode is set to ON. If the Span Factor is set to OFF,
the adjustment range will not appear.
13. Disconnect the tubing from the EAGLE 2’s probe.
14. Repeat Step 6 through Step 13 for any other channels you want to calibrate. Make
sure you use an appropriate calibration cylinder for each sensor.
CAUTION:
When calibrating the oxygen sensor, verify the concentration of oxygen
listed on the cylinder’s label. For oxygen-free samples (100% nitrogen for
example), set the oxygen calibration value to 0.0%.
15. After the last channel is calibrated, disconnect the calibration tubing from the probe,
then unscrew the demand flow regulator from the calibration cylinder.
16. With the Select Sensor Screen displayed, place the cursor next to
ESCAPE
using the
AIR
▲
YES button.
17. Press and release the POWER ENTER RESET button to return to the Calibration
Mode Screen.
18. Use the RANGE
▼
SHIFT button to place the cursor next to the
NORMAL
OPERATION
menu item, then press and release the POWER ENTER RESET
button to return to Measuring Mode.
SELECT SENSOR
TO CALIBRATE
>ESCAPE
CH4 OXY
H2S CO