72 • Replacing the Hydrophobic Probe’s Particle Filter and Hydrophobic Filter
Replacing the Hydrophobic Probe’s
Particle Filter and Hydrophobic Filter Disk
Inspect the probe’s internal components if you notice that the EAGLE 2’s pump sounds
bogged down or if an unexplained low flow alarm occurs. Replace the particle filter if it
appears to be dirty. Replace the hydrophobic filter disk if it appears dirty or saturated with
liquid. Replace the O-rings in the probe if either of them appears damaged.
1. Grasp each end of the clear probe body firmly and unscrew the two halves from each
other. One half includes a plastic tube fitting and the probe tube. The other half
includes a metal fitting that mates with the EAGLE 2 inlet fitting.
Figure 21: Replacing the Particle Filter and Hydrophobic Filter Disk
2. Remove the white hydrophobic filter disk from the top of the particle filter or from
the probe body.
3. Remove the particle filter from the probe body.
Probe Body
Particle Filter, Not Included
in PID Probe
O-ring (Inside)
Hydrophobic Filter Disk
O-ring