74 • Replacing the Hydrophobic Filter
EAGLE 2 Operator’s Manual
6. Locate the hydrophobic filter. It is over the oxygen sensor. Note which side of the
hydrophobic filter has the RKI logo and part number. This is the inlet side and should
be facing toward the front of the EAGLE 2.
Figure 22: Replacing the Hydrophobic Filter
7. Pull the grey connecting tubes off of each end of the filter and remove it.
8. Install the new filter with the red RKI logo and part number on the inlet side of the
flow chamber, facing the front of the EAGLE 2. Make sure to push the grey
connection tubes all the way onto the filter’s hose barbs.
9. Confirm that the main PCB is seated in its slots and that its bottom edge is resting on
the bottom of the bottom case. If the main PCB is not seated properly, then it may be
damaged when the top case is re-installed.
10. Make sure that the top case gasket is fully seated in its groove and carefully put the
top case back on the bottom case. If you have any difficulty mating the top and
bottom cases, inspect the placement of the main PCB and the placement of the top
case gasket.
11. Turn the EAGLE 2 upside down and tighten the three case screws to secure the top
case to the bottom case.
Grey Connecting
Tube, 2X
Hydrophobic Filter
RKI Logo & Part Number on This Side
Bottom Case Shown Without
Tubing For Clarity