EAGLE 2 Operator’s Manual
Calibrating Using the Single Calibration Method • 91
4. Move the cursor next to the
SINGLE CALIBRATION
menu item by using the AIR
▲
YES button.
5. Press and release the POWER ENTER RESET button. The Select Sensor Screen
appears with the cursor flashing.
6. Move the cursor next to the sensor you want to calibrate with the AIR
▲
YES and
RANGE
▼
SHIFT buttons. In the example below, the CH
4
sensor is selected for span
adjustment.
If you do not want to proceed with the span adjustment, press and release the
DISPLAY ADJUST NO button or place the cursor next to
ESCAPE
and press and
release POWER ENTER RESET to return to the Calibration Mode Screen.
If you do want to continue with the calibration, proceed with the next step.
7. Press and release the POWER ENTER RESET button to proceed to the Single
Calibration Gas Value Screen for the selected channel. The calibration gas value is
flashing.
CALIBRATION MODE
AUTO CALIBRATION
> SINGLE CALIBRATION
PERFORM AIR ADJUST
NORMAL OPERATION
SELECT SENSOR
TO CALIBRATE
>ESCAPE
CH4 OXY
H2S CO
SELECT SENSOR
TO CALIBRATE
ESCAPE
>CH4 OXY
H2S CO
SINGLE CALIBRATION
CH4 50 %LEL
UP/DOWN TO ADJUST
CALIBRATION VALUE
ENTER WHEN DONE