GX-3R Pro Operator’s Manual
Chapter 5: User Mode and Calibration
•
84
Setting the Calibration Values in SETTING CAL--P
1. While in User Mode, use AIR to place the cursor next to
GAS CAL
.
2. Press and release POWER MODE. The Gas Cal menu appears.
3. Use AIR to place the cursor next to
AUTO CAL
.
4. Press and release POWER MODE. The Auto Cal screen for gases assigned to Cylinder A
displays.
5. Use AIR to scroll to
SETTING CAL--P
.
6. Press and release POWER MODE. The combustible gas channel displays.
MAINTE
7:49
BUMP TEST
GAS CAL
CAL SETTING
USER MODE
MAINTE
7:49
CO2 ZERO CAL
AUTO CAL
ESCAPE
GAS CAL
MAINTE
7:49
SETTING CAL--P
MAINTE
- - - -
CH4 %LEL
SETTING CAL--P