GX-3R Pro Operator’s Manual
Chapter 5: User Mode and Calibration
•
90
CAL REMINDER
ON
(factory setting): The GX-3R Pro gives an indication at start up if it is due for calibration. The
type of indication depends on the
CAL EXPIRED
setting (see page 91).
OFF
: The GX-3R Pro does not give an indication at start up if it is due for calibration.
1. While in the
CAL SETTING
menu, use AIR to place the cursor next to
CAL REMINDER
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
CAL SETTING
menu.
5. See “ESCAPE” on page 93 to return to User Mode.
CAL INTERVAL
This setting defines the amount of time between calibrations. The time can be set in 1 day
increments. The minimum setting is
1
day and the maximum setting is
1000
days. The factory
setting is
90
days.
1. While in the
CAL SETTING
menu, use AIR to place the cursor next to
CAL INTERVAL
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
CAL SETTING
menu.
5. See “ESCAPE” on page 93 to return to User Mode.
MAINTE
7:49
CAL REMINDER
CAL INTERVAL
CAL EXPIRED
CAL SETTING
MAINTE
7:49
CAL REMINDER
CAL INTERVAL
CAL EXPIRED
CAL SETTING