M2A Transmitter Operator’s Manual
43
Table 11: Configuration Parameters
Parameter
(Factory Set Value)
Description
ALARM-1 (level)
(See Table 1 on page 7)
The gas reading at which the M2A initiates an alarm 1 condition.
ALARM-1 (activation)
(Decrease for oxygen M2A, Increase for
all other types)
Indicates if the alarm 1 circuit is activated by gas readings increasing (
Increase
)
or decreasing (
Decrease
) to the ALARM-1 Level.
ALARM-1 (relay action)
(N. DE-EN)
If set as
N. DE-EN
, the alarm 1 relay is de-energized in normal operation and
energizes when an alarm 1 condition is initiated.
If set as
N. EN
, the alarm 1 relay is energized in normal operation and de-ener
-
gizes when an alarm 1 condition is initiated.
ALARM-1 (relay reset)
(LATCH)
If set as
LATCH
, you must press the ENTER button to reset the alarm 1 circuit
after the alarm 1 condition passes.
If set as
SELF-RST
, the M2A automatically resets the alarm 1 circuit after the
alarm 1 condition passes.
A1 OnDy (alarm 1 on delay)
(1 secs)
The amount of time the M2A delays activation of the alarm 1 circuit once an
alarm 1 condition is initiated. It can be set in 1 second increments from 0 - 60
seconds, in 1 minute increments from 1 - 15 minutes, and in 15 minute incre
-
ments from 15 - 60 minutes.
A1 OffDy (alarm 1 off delay)
(0 sec.)
The amount of time the M2A delays turning off the alarm 1 circuit once an alarm
1 condition passes. This parameter appears only if the alarm 1 relay reset setting
is set to
SELF-RST
. It can be set in 1 second increments from 0 - 60 seconds, in
1 minute increments from 1 - 15 minutes, and in 15 minute increments from 15 -
60 minutes.
ALARM-2 (level)
(See Table 1 on page 7)
The gas reading at which the M2A initiates an alarm 2 condition.
ALARM-2 (activation)
(Increase for all types)
Indicates if the alarm 2 circuit is activated by gas readings increasing (
Increase
)
or decreasing (
Decrease
) to the ALARM-1 Level.
ALARM-2 (relay action)
(N. DE-EN)
If set as
N. DE-EN
, the alarm 2 relay is de-energized in normal operation and
energizes when an alarm 2 condition is initiated.
If set as
N. EN
, the alarm 2 relay is energized in normal operation and de-ener
-
gizes when an alarm 2 condition is initiated.
ALARM-2 (relay reset)
(LATCH)
If set as
LATCH
, you must press the ENTER button to reset the alarm 2 circuit
after the alarm 2 condition passes.
If set as
SELF-RST
, the M2A automatically resets the alarm 2 circuit after the
alarm 2 condition passes.
A2 OnDy (alarm 2 on delay)
(1 secs)
The amount of time the M2A delays activation of the alarm 2 circuit once an
alarm 2 condition is initiated. It can be set in 1 second increments from 0 - 60
seconds, in 1 minute increments from 1 - 15 minutes, and in 15 minute incre
-
ments from 15 - 60 minutes.
A2 OffDy (alarm 2 off delay)
(0 sec.)
The amount of time the M2A delays turning off the alarm 2 circuit once an alarm
2 condition passes. This parameter appears only if the alarm 2 relay reset setting
is set to
SELF-RST
. It can be set in 1 second increments from 0 - 60 seconds, in
1 minute increments from 1 - 15 minutes, and in 15 minute increments from 15 -
60 minutes.