OX-04 Operator’s Manual
Chapter 5: User Mode and Calibration
•
55
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
CAL.RMDR
item in
the
CAL SET
menu.
5. See “Exiting the CAL SET Menu” on page 56 to return to User Mode.
CAL.INT
This setting defines the amount of time between calibrations and can be set in 1 day increments.
The minimum setting is
1
day and the maximum setting is
1000
days. The factory setting is
90
days.
1. After entering the
CAL SET
menu, press AIR to scroll to
CAL.INT
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
CAL.INT
item in the
CAL SET
menu.
5. See “Exiting the CAL SET Menu” on page 56 to return to User Mode.
CAL.EXPD
This item defines what indicator is given during start up when calibration is due and
CAL.RMDR
is set to
ON
(factory setting).
CONFIRM
(factory setting): The OX-04 gives an indicator at start up if calibration is past due.
Press and release AIR to continue without calibrating or press and release POWER MODE to
enter User Mode and perform a calibration.
CANT.USE
: The OX-04 gives an indicator at start up that calibration is past due. Press and
release POWER MODE to enter User Mode and perform a calibration. Pressing AIR has no
effect. A successful calibration must be performed in order to use the instrument.
NONE
: The OX-04 gives an indicator at startup that calibration is past due. It is not necessary to
acknowledge the indicator. If desired, press POWER MODE to enter User Mode and perform a
calibration. The warm-up sequence continues on its own.
1. After entering the
CAL SET
menu, press AIR to scroll to
CAL.EXPD
.
CAL .INT
CAL .EXPD