Setting the Operational Parameters in Edit Mode • 21
5. Use the BUMP
and CAL
buttons to scroll through the different calibration
parameters. Do not scroll past the GAS--IN parameter until you have finished
making any desired changes.
6. To edit a parameter, press and release the EDIT ENTER button when that
parameter is displayed. An “E” will appear to the right of the parameter name
indicating that the parameter value can be changed. In the example below, the
AIR--IN parameter has been selected for updating.
7. Use the BUMP
and CAL
buttons to set the parameter to the desired
value, then press and release the EDIT ENTER button. The “E” to the right of
parameter name will disappear.
8. Repeat step 6 - step 7 to set any other parameters.
9. When you are done setting the parameters, use the BUMP
button to scroll
past the GAS--IN parameter screen. The screen will indicate that the
parameter changes have been saved.
The instrument will return to the home screen.
45
AIR--IN
90
GAS--IN
45 90
DECISION
50 12.0
TRANSMIT
CO ppm H2S ppm
CH4 %LEL OXY vol%
50 25.0
45
E AIR--IN
30 90
DECISION