Chapter 9: Viewing Data in the Logs Window • 121
5.
To view the interval trend data, double click the desired file.
Figure 72: Interval Trend Data
a. To change the layout of the table and chart, click any of the four layout control buttons in
the upper left corner of the screen.
b. To change the proportions of the table to the graph, click the divider between the two and
move it.
c. To display instrument information along with the data and graph, click the “i” control
button in the upper left corner.
6.
In the table part of the screen:
a. The interval trend data is shown as the average gas readings over the user defined
interval trend time. So if the data logging session started at 4:13:38 PM and the interval
time is set to 1 minute, then the readings logged at 4:14:38 PM are the average reading
for each channel over that one minute period. If a channel went into Warning or Alarm, it
will be indicated.
b. Events are displayed whether they occurred at scheduled log times or in between them.
Events include gas alarms such as a warning condition, trouble conditions such as a
sensor failure, and an indication that the GX-3R (Pro) is returning to “normal” condition
after an alarm has been reset using the POWER MODE button.
Layout Control Buttons Print Save Export Restore Previous