Chapter 7: Changing GX-3R (Pro) Parameters • 83
Interval Time (Sec)
The interval time is how often the GX-3R (Pro) will log interval trend readings. It can be set to
10
,
20
,
30
(factory setting),
60
,
180
,
300
, or
600
seconds. To change the interval time, use the arrows to the
right of the current value to increase or decrease it. You may also highlight the current setting and type
in the new desired setting. The value that you type in must match the allowable interval settings. If it
does not, when you save your changes, your entered value will be shifted to the closest match. For
example, if you enter 90 seconds, when you save the changes, the interval will adjust itself to
60
seconds.
Memo Section
The Memo section allows the user to enter remarks for a GX-3R (Pro) that is connected to the
program. The Memo may be used to document anything of interest, such as maintenance records
(noting when a sensor or a battery pack was changed). The Memo section is empty until the user enters
information.
The first time information is entered, a Memo file is created in the Logs window. If the Memo is
changed in the same operating session, that Memo file is updated. When the GX-3R (Pro) is
disconnected from the PC Program and connected again, a new Memo file is created with the current
memo. If you edit the memo field in that operating session, the new Memo file is updated. If you clear
the Memo field for a GX-3R (Pro), the Memo file that was created when the GX-3R (Pro) was
connected to the PC Program will be deleted since it is now empty. All memos from previous operating
sessions will always be available in the Logs window. They cannot be deleted.
To enter information into the Memo section,
1.
Place the cursor in the Memo field. If a memo has been previously entered, it will still appear
in the Memo field.
2.
If necessary, delete the old memo. Enter the information you wish to document.
3.
Click